From Wikipedia, the free encyclopedia

Software installed in medical devices is assessed for health and safety issues according to international standards.

Safety classes

[edit]

Software classification is based on the potential for hazard(s) that could cause injury to the user or patient.[1]

Per IEC 62304:2006 + A1:2015, the software can be divided into three separate classes: class A, B, and C.

  • The SOFTWARE SYSTEM is software safety class A if:
    • the SOFTWARE SYSTEM cannot contribute to a HAZARDOUS SITUATION; or
    • the SOFTWARE SYSTEM can contribute to a HAZARDOUS SITUATION which does not result in unacceptable RISK after consideration of RISK CONTROL measures external to the SOFTWARE SYSTEM.
  • The SOFTWARE SYSTEM is software safety class B if:
    • the SOFTWARE SYSTEM can contribute to a HAZARDOUS SITUATION which results in unacceptable RISK after consideration of RISK CONTROL measures external to the SOFTWARE SYSTEM and the resulting possible HARM is non-SERIOUS INJURY.
  • The SOFTWARE SYSTEM is software safety class C if:
    • the SOFTWARE SYSTEM can contribute to a HAZARDOUS SITUATION which results in unacceptable RISK after consideration of RISK CONTROL measures external to the SOFTWARE SYSTEM and the resulting possible HARM is death or SERIOUS INJURY“

Serious injury

[edit]

For the purpose of this classification, serious injury is defined as injury or illness that directly or indirectly is life-threatening; results in permanent impairment of a body function or permanent damage to a body structure; or necessitates medical or surgical intervention to prevent permanent impairment of a body function or permanent damage to a body structure.